Abstract Despite the widespread use of narrow diameter inflatable penile prosthesis (IPP) cylinders in patients with corporal fibrosis, outcomes data is sparse. We evaluated patients who underwent IPP placement with AMS™ 700 controlled expansion restricted (CXR) cylinders from 2007-2021. Patient characteristics, device details, and surgical outcomes were assessed. A non-validated questionnaire was also distributed to patients to assess satisfaction. Among 982 IPPs placed over the study period at our institution, 49 (5.3%) used CXR cylinders. Indications for narrow cylinders: prior explant for infection (67.3%), ischemic priapism (16.3%), and idiopathic fibrosis (16.3%). Median corporal length was 19cm (IQR 17-21cm). Discordant intraoperative corporal measurements (24.5%) and rear tip extender lengths (26.5%) were common. Post-operative complications occurred in 8 patients (16.3%) and included 3 infections (6.1%), 2 cylinder herniations (4.1%), 2 mechanical failures (4.1%) and one case of glans necrosis (2.0%). Five of these (10.2%) required explantation, while 4 (8.2%) were managed with revision. Over the follow-up period, 73% of patients reported satisfaction with rigidity. Primary drivers of dissatisfaction were perceived loss of penile length and girth. The AMS™ 700 CXR is a useful tool for challenging corporal fibrosis cases and shows acceptable surgical outcomes with moderate patient acceptability.