Does pulsed-field ablation improve efficacy and safety compared to radiofrequency or cryoballoon ablation in patients with atrial fibrillation?
Pulsed-field ablation is associated with shorter procedure times and lower rates of atrial arrhythmia recurrence compared to radiofrequency ablation, as well as lower recurrence and complication rates compared to cryoballoon ablation.
This review investigated efficacy and safety of pulsed-field ablation (PFA) in comparison with radiofrequency ablation (RFA), cryoballoon ablation (CBA), or both combined. The Odds ratio (OR) and mean difference (MD) with 95% confidence interval (95% CI) were computed. PFA allowed shorter procedure (MD -44.27 minutes, 95% CI: -63.61; -24.93) and left atrium (LA) dwell (MD -32.71 minutes (95% CI: -58.64; -6.78) times, but with longer fluoroscopy time than RFA (MD 8.54 minutes, 95% CI: 4.03; 13.04). Post-procedural complications rate was lower with PFA than CBA (OR 0.53, 95% CI: 0.35, 0.80). Atrial arrhythmias recurrence rate within one year of follow-up was lower with PFA than RFA (OR 0.68, 95% CI; 0.53; 0.87) and CBA (OR 0.69, 95% CI: 0.48; 0.97). PFA allowed shorter procedure and LA dwell times, as well as lower atrial arrhythmia recurrence than RFA and lower post-procedural complications and atrial arrhythmias recurrence rates than CBA.