Micra leadless pacemakers were associated with lower odds of complications compared with transvenous pacemakers at up to 1 year (3.30% vs 7.43%; OR 0.49, 95% CI 0.34-0.70).
Meta-Analysis
Do leadless pacemakers reduce complications and provide acceptable pacing capture thresholds compared to transvenous pacemakers?
Leadless pacemakers, particularly the Micra device, demonstrate a low risk of complications and good electrical performance up to 1 year, with lower complication odds than transvenous pacemakers.
Effect estimate: OR 0.49 (95% CI 0.34-0.70)
Absolute Event Rate: 3.3% vs 7.43%
Background Leadless pacemaker is a novel technology, and evidence supporting its use is uncertain. We performed a systematic review and meta-analysis to examine the safety and efficacy of leadless pacemakers implanted in the right ventricle. Methods and Results We searched PubMed and Embase for studies published before June 6, 2020. The primary safety outcome was major complications, whereas the primary efficacy end point was acceptable pacing capture threshold (≤2 V). Pooled estimates were calculated using the Freedman-Tukey double arcsine transformation. Of 1281 records screened, we identified 36 observational studies of Nanostim and Micra leadless pacemakers, with most (69.4%) reporting outcomes for the Micra. For Micra, the pooled incidence of complications at 90 days (n=1608) was 0.46% (95% CI, 0.08%-1.05%) and at 1 year (n=3194) was 1.77% (95% CI, 0.76%-3.07%). In 5 studies with up to 1-year follow-up, Micra was associated with 51% lower odds of complications compared with transvenous pacemakers (3.30% versus 7.43%; odds ratio OR, 0.49; 95% CI, 0.34-0.70). At 1 year, 98.96% (95% CI, 97.26%-99.94%) of 1376 patients implanted with Micra had good pacing capture thresholds. For Nanostim, the reported complication incidence ranged from 6.06% to 23.54% at 90 days and 5.33% to 6.67% at 1 year, with 90% to 100% having good pacing capture thresholds at 1 year (pooled result not estimated because of the low number of studies). Conclusions Most studies report outcomes for the Micra, which is associated with a low risk of complications and good electrical performance up to 1-year after implantation. Further data from randomized controlled trials are needed to support the widespread adoption of these devices in clinical practice.
Ngo et al. (Fri,) conducted a meta-analysis in Indication for right ventricular pacing. Leadless pacemakers (Micra and Nanostim) vs. Transvenous pacemakers was evaluated on Major complications (OR 0.49, 95% CI 0.34-0.70). Micra leadless pacemakers were associated with lower odds of complications compared with transvenous pacemakers at up to 1 year (3.30% vs 7.43%; OR 0.49, 95% CI 0.34-0.70).
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: