July 24, 2014IEEE Transactions on Biomedical Engineering139 citationsOpen Access

Remote Detection of Photoplethysmographic Systolic and Diastolic Peaks Using a Digital Camera

View Full Paper
DMDaniel McDuffSGSarah GontarekRPRosalind W. Picard

Structured PICO

Does remote photoplethysmogram measurement using a digital camera accurately detect systolic and diastolic peaks compared to a contact fingertip sensor?

I
Intervention
Remote photoplethysmogram signal measurement using a digital camera (orange, green, and cyan color channels) from a distance of 3 m
C
Comparator
Contact fingertip sensor
O
Outcome
Agreement in peak and interval timings (interbeat intervals and systolic-diastolic peak-to-peak times)surrogate

Remote photoplethysmography using a digital camera can accurately measure interbeat intervals and systolic-diastolic peak-to-peak times compared to contact sensors.

Abstract

We present a new method for measuring photoplethysmogram signals remotely using ambient light and a digital camera that allows for accurate recovery of the waveform morphology (from a distance of 3 m). In particular, we show that the peak-to-peak time between the systolic peak and diastolic peak/inflection can be automatically recovered using the second-order derivative of the remotely measured waveform. We compare measurements from the face with those captured using a contact fingertip sensor and show high agreement in peak and interval timings. Furthermore, we show that results can be significantly improved using orange, green, and cyan color channels compared to the tradition red, green, and blue channel combination. The absolute error in interbeat intervals was 26 ms and the absolute error in mean systolic-diastolic peak-to-peak times was 12 ms. The mean systolic-diastolic peak-to-peak times measured using the contact sensor and the camera were highly correlated, ρ = 0.94 (p 0.001). The results were obtained with a camera frame-rate of only 30 Hz. This technology has significant potential for advancing healthcare.

Ask AI
Helpful
Bookmark
Share
View Full Paper

Cite This Study

McDuff et al. (2014) studied this question.

synapsesocial.com/papers/6a1a3dc7707b23dfc4ef13adhttps://doi.org/10.1109/tbme.2014.2340991