Purpose To present a new complex‐valued B 1 + mapping method for electrical properties tomography using Carr‐Purcell spin echoes. Methods A Carr‐Purcell (CP) echo train generates pronounced flip‐angle dependent oscillations that can be used to estimate the magnitude of B 1 + . To this end, a dictionary is used that takes into account the slice profile as well as T 2 relaxation along the echo train. For validation, the retrieved B 1 + map is compared with the actual flip angle imaging (AFI) method in a phantom (79 ε 0 , 0.34 S/m). Moreover, the phase of the first echo reflects the transceive phase. Overall, the CP echo train yields an estimate of the complex‐valued B 1 + , allowing electrical properties tomography with both permittivity and conductivity. The presented method is evaluated in phantom scans as well as for in vivo brain at 3 T. Results In the phantom, the obtained magnitude B 1 + maps retrieved from the CP echo train and the AFI method show excellent agreement, and both the reconstructed estimated permittivity (79 ± 3) ε 0 and conductivity (0.35 ± 0.04) S/m values are in accordance with expectations. In the brain, the obtained electrical properties are also close to expectations. In addition to the retrieved complex B 1 + information, the decay of the CP echo trains also yields an estimate for T 2 . Conclusion The CP sequence can be used to simultaneously provide both B 1 + magnitude and phase estimations, and therefore allows for full reconstruction of the electrical properties.
Iyyakkunnel et al. (Tue,) studied this question.