Fourier domain optical coherence tomography demonstrated excellent reproducibility for consecutive pullbacks in stented coronaries, with low inter-software variability (R2=0.98-1.00, p<0.01).
Observational
Does Fourier domain optical coherence tomography (FD OCT) provide reproducible quantitative measurements of in vivo stented coronary arteries across different software packages?
FD OCT demonstrates excellent reproducibility for consecutive pullbacks and reliable quantitative analysis across different software packages in stented coronary arteries.
Effect estimate: R2=0.98 ~ 1.00
p-value: p=<0.01
AIMS: Fourier domain optical coherence tomography (FD OCT) enables imaging of long coronary artery segments within few seconds, employing high data acquisitions, speed and fast automated catheter pullback. However, the reproducibility of these high-speed pullbacks in the clinical situation is unknown. We tested the reproducibility of in vivo, intracoronary FD OCT and assessed the influence of different computer-assisted algorithms on quantitative analysis. METHODS AND RESULTS: In patients undergoing elective coronary stenting, two repeated FD OCT pullbacks (20 mm/sec), were acquired. Lumen area (LA) and stent area (SA) were measured at 1 mm longitudinal intervals (n=18 pullbacks, n=326 frames). Inter-study variability in terms of absolute difference of mean LA, mean SA and minimum LA was very low (-0.06±0.28 mm2, -0.05±0.29 mm2 and -0.11±0.33 mm2 in software 1) Sources of variability were incomplete visualisation of the vessel circumference, ambiguous luminal borders and drift of internal catheter calibration (Z-offset). Inter-software variability for LA and SA was low (R2=0.98 ~ 1.00, p<0.01, respectively). CONCLUSIONS: FD OCT shows excellent reproducibility for consecutive pullbacks and represents a reliable tool for the in vivo assessment of stented coronaries. Computer-assisted quantitative analysis of FD OCT may be a valuable tool for future studies.
أجرى أوكامورا وزملاؤه (Sun) دراسة استطلاعية في تركيب الدعامات التاجية الانتقائية. تم تقييم تقنية تصوير التداخل البصري المدى الترددي (FD OCT) مقابل السحب المتكررة وحزم البرمجيات المختلفة على تباين متوسط مساحة اللمعة، ومتوسط مساحة الدعامة، وأقل مساحة لللمعة (R2=0.98 ~ 1.00, p=<0.01). أظهرت تقنية تصوير التداخل البصري المدى الترددي تكرارية ممتازة للسحوبات المتتالية في الشرايين التاجية المثبتة، مع تباين منخفض بين البرمجيات (R2=0.98-1.00, p<0.01).
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: