The aim of this in vitro study was to analyze the impact of model aging on the accuracy of 3D-printed dental models. A maxillary full-arch reference model with prepared teeth for a three-unit fixed dental prosthesis was scanned ten times with an intraoral scanner (3Shape TRIOS Pod) and ten models were 3D printed (Straumann P-Series). All models were stored under constant conditions and digitized with a desktop scanner after 1 day; 1 week; and 2, 3, and 4 weeks. For accuracy, a best-fit algorithm was used to analyze the deviations of the abutment teeth (GFaI e.V Final Surface®). Wilcoxon Rank Sum Tests were used for comparisons with the level of significance set at α = 0.05. Deviation analysis of the tested models showed homogenous intragroup distance calculations at each timepoint. The most accurate result was for 1 day of aging (3.3 ± 1.3 µm). A continuous decrease in accuracy was observed with each aging stage from day 1 to week 4. A time-dependent difference was statistically significant after 3 weeks (p = 0.0008) and 4 weeks (p < 0.0001). Based on these findings, dental models should not be used longer than 3 to 4 weeks after 3D printing for the fabrication of definitive prosthetic reconstructions.