Does the Nexfin device accurately measure rapid changes in cardiac index and predict fluid responsiveness compared to transpulmonary thermodilution in patients following cardiac surgery?
The non-invasive Nexfin device is not sufficiently accurate to measure rapid changes in cardiac index or predict fluid responsiveness in post-cardiac surgery patients compared to transpulmonary thermodilution.
BACKGROUND: The Nexfin device uses non-invasive photoplethysmography to monitor cardiac output and respiratory variations in pulse pressure and stroke volume. The aim of this study was to compare rapid changes in cardiac index after fluid challenge between Nexfin and bolus transpulmonary thermodilution and the ability to predict fluid responsiveness of dynamic indices given by Nexfin. METHODS: Simultaneous comparative cardiac index were collected from transpulmonary thermodilution and Nexfin before and after fluid challenge in 45 patients following conventional cardiac surgery. Correlations, Bland-Altman analyses and percentage errors were calculated. Pulse pressure variations and stroke volume variations before fluid challenge were collected to assess their discrimination in predicting fluid responsiveness. RESULTS: Eight (18%) patients were excluded. A weak positive relationship was found between rapid changes in cardiac index after fluid challenge given by both technologies (n = 37, r = 0.39, P = 0.019). Bias, precision and limits of agreements were 0.20 l/min/m(2) (95% confidence interval (CI) 0.02-0.40), 0.57 l/min/m(2) and ± 1.12 l/min/m(2) before fluid challenge, and 0.01 l/min/m(2) (95% CI -0.24 to 0.26), 0.74 l/min/m(2) and ± 1.45 l/min/m(2) after fluid challenge. Percentage errors between Nexfin and transpulmonary thermodilution were 55% and 58% before and after fluid challenge, respectively. Pulse pressure variations and stroke volume variations given by Nexfin were not discriminant to predict fluid responsiveness: areas under receiver operating characteristics curves 0.57 (95% CI 0.40-0.73) and 0.50 (0.33-0.67), respectively. CONCLUSIONS: The Nexfin cannot be used to measure rapid changes in cardiac index following fluid challenge and to predict fluid responsiveness after cardiac surgery.
Fischer et al. (2013) studied this question.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: