The epicardial Pulsed Field Ablation clamp reliably produced transmural lesions in a vegetal model using 2 or 4 applications, demonstrating a linear relationship between lesion volume and voltage.
Does a novel multielectrode surgical PFA clamp produce controlled transmural lesions in an ex-vivo vegetal model?
A novel multielectrode surgical PFA clamp demonstrated the ability to create controlled, transmural lesions in an ex-vivo vegetal model, showing potential for consistent ablation.
Absolute Event Rate: 0% vs 0%
The epicardial PFA clamp reliably produced transmural lesions in a vegetal model of TS using 2 or 4 applications. With 4 applications, lesion volume demonstrated a predictable, linear relationship with increasing voltage, supporting the device's potential for controlled and consistent ablation.
Valappil et al. (Sat,) reported a other. The epicardial Pulsed Field Ablation clamp reliably produced transmural lesions in a vegetal model using 2 or 4 applications, demonstrating a linear relationship between lesion volume and voltage.