Is the Atrial High Rate Episode Diagnostic Feature Reliahle in Detecting Paroxysmal Episodes of Atrial Tachyarrhythmias?

View Full Paper
KSK. SeidlEMECKARD MEISELEVEddy Vanagt

Key Result

The Thera pacemaker diagnostic algorithm demonstrated highly reliable detection of atrial tachyarrhythmias, achieving 98% sensitivity and 100% specificity when programmed with optimized settings.

Key Points

  • To evaluate the accuracy and reliability of the Thera pacemaker's atrial high rate episode diagnostic feature in detecting paroxysmal atrial tachyarrhythmias.
  • Compared pacemaker-detected atrial tachyarrhythmia episodes against 45 ambulatory Holter recordings (30 showing sinus rhythm, 15 showing tachyarrhythmias).
  • Evaluated default settings (detection rate 160 beats/min, detection beats 40, termination beats 10) and performed software simulations with alternative programming parameters.
  • Default programming caused 12 false-positive episodes across 4 of 30 sinus rhythm recordings due to far-field R- and T-wave oversensing, and missed 32 of 125 arrhythmia episodes (93 detected) because of short duration.
  • Software simulation with modified parameters (detection rate 220 beats/min, detection beats 10, termination beats 20) demonstrated 98% sensitivity and 100% specificity for detecting atrial tachyarrhythmias.

Study Design

Type

Observational (n=45)

Structured PICO

Does the Thera pacemaker atrial high rate episode diagnostic feature reliably detect paroxysmal episodes of atrial tachyarrhythmias compared to Holter monitoring?

P
Population
45 ambulatory (Holter) recordings
I
Intervention
Thera pacemaker atrial high rate episode diagnostic feature
C
Comparator
Holter monitoring
O
Outcome
Detection of episodes of atrial fibrillation/tachyarrhythmiassurrogate

The Thera pacemaker's high rate episode monitor is a reliable tool for detecting atrial tachyarrhythmias when programmed with optimized settings (detection rate 220 beats/min, detection beats 10, termination beats 20).

Abstract

The atrial high rate episode diagnostic in the Thera pacemaker reports frequency, duration, and date/time of atrial tachyarrhythmias according to programmed criteria. The aim of the study was to validate the atrial high rate episode diagnostic feature. Episodes of atrial fibrillation recorded by Holter monitoring were compared to episodes detected by the pacemaker. Forty five ambulatory (Holter) recordings were used for evaluation. Thirty of 45 ambulatory (Holter) recordings showed sinus rhythm. On 4 of these 30 ambulatory (Holter) recordings, the Thera detected 12 episodes of atrial tachyarrhythmias as false-positives (sinus rhythm was detected as atrial tachyarrhythmia). The main reason was far-field R and T was oversensing. On 15 of 45 ambulatory (Holter) recordings, 125 episodes of atrial tachyarrhythmias were recorded. Ninety-three of these events also were detected by the pacemaker, while for 32 events the Thera reported sinus rhythm. The main reason was that the episodes were of too short duration. Therefore, the Thera (programmed with detection rate 160 beats/min, detection beats 40, termination beats 10) was unable to detect atrial tachycardias. Software simulation of the diagnostic algorithm under several programming settings using the digitized Holter files demonstrated highly reliable detection of atrial tachyarrhythmias (sensitivity 98%, specificity 100%) when programmed as follows: detection rate 220 beats/min, detection beats 10, termination beats 20. It can be concluded that Thera's high rate episode monitor is a reliable tool for detection of atrial tachyarrhythmias, if programmed as recommended.

Ask AI
Helpful
Bookmark
Share
View Full Paper

Cite This Study

Seidl et al. (1998) conducted an observational in Paroxysmal atrial tachyarrhythmias (n=45). Thera pacemaker atrial high rate episode diagnostic feature vs. Holter monitoring was evaluated on Detection of atrial tachyarrhythmias (sensitivity and specificity). The Thera pacemaker diagnostic algorithm demonstrated highly reliable detection of atrial tachyarrhythmias, achieving 98% sensitivity and 100% specificity when programmed with optimized settings.

synapsesocial.com/papers/6a120c84a2d24b27c166907ahttps://doi.org/10.1111/j.1540-8159.1998.tb00125.x