December 1, 1967Circulation Research27 citations

An Electromagnetic Catheter-Flowmeter

View Full Paper
AKAlexander KolinJAJames D. ArcherGRGordon Ross

Key Result

An electromagnetic catheter-flowmeter successfully measured phasic blood flow in the left renal and superior mesenteric arteries of a dog.

PICO

I
Intervention / Comparator
Electromagnetic catheter-flowmeter vs Calibrated noncannulating electromagnetic flow transducer
O
Primary Outcome
Phasic blood flow measurement

Abstract

An electromagnetic flowmeter incorporated into a catheter is described. The catheter is introduced into the aorta via the femoral artery. The flow sensor is located some distance below the catheter tip and is sensitive to flow at right angles to the catheter. A crater-like elevation at one end of the transducer lumen permits funnel-like connection of the transducer to the ostium of an aortic branch. The transducer thus measures the blood flow through a selected branch of the aorta. It is pressed against the branch artery by arching the catheter by means of a pull-wire. A reliable and readily reproducible zero-flow baseline is obtained by slightly withdrawing the lumen of the transducer from the branch ostium, thus closing the lumen by the aortic wall. Records of phasic blood flow in the left renal and superior mesenteric arteries of a dog are presented. The calibration was accomplished by comparison with a calibrated noncannulating electromagnetic flow transducer. Reliable calibration can also be performed in vitro.

Ask AI
Helpful
Bookmark
Share
View Full Paper

Cite This Study

Kolin et al. (1967) studied this question. Electromagnetic catheter-flowmeter vs. Calibrated noncannulating electromagnetic flow transducer was evaluated on Phasic blood flow measurement. An electromagnetic catheter-flowmeter successfully measured phasic blood flow in the left renal and superior mesenteric arteries of a dog.

synapsesocial.com/papers/6a1eab39ae66660099a42a50https://doi.org/10.1161/01.res.21.6.889
Ask AI
Helpful
Bookmark
Share
View Full Paper

Also Consider

Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context:

  1. 1The external field intravascular electromagnetic flowmeter system as applied to standard arteriographic catheters and conscious humans1976 · 5 citations
  2. 2Evaluation of a new catheter‐mounted electromagnetic velocity sensor during cardiac catheterization1980 · 29 citations
  3. 3Approaches to blood-flow measurement by means of electromagnetic catheter flow meters1970 · 9 citations
  4. 4A Newly Modified Electromagnetic Blood Flowmeter Capable of High Fidelity Flow Registration1952 · 28 citations
  5. 5A Magnetic Flowmeter for Recording Cardiac Output1959 · 4 citations