January 1, 1984Japanese Circulation Journal-english Edition57 citationsOpen Access

A very sensitive direct radioimmunoassay system for plasma angiotensin II and its clinical application in various hypertensive diseases.

KSKazuaki ShimamotoHIHiroyuki IshidaYNYasuyuki Nakahashi

Key Points

  • To develop a highly sensitive direct radioimmunoassay for plasma angiotensin II that bypasses extraction procedures and evaluate its diagnostic utility in hypertensive diseases.
  • Developed an unextracted radioimmunoassay using 100 µL of plasma, specific rabbit antiserum against synthetic angiotensin II (1:1,500,000 dilution), and 125I-labeled angiotensin II (specific activity: 1,600 µCi/µg).
  • Evaluated assay validity through recovery tests, standard curve parallelism, cross-reactivity with angiotensin I and III, and correlation against standard Dowex column extraction.
  • Measured fasting plasma angiotensin II levels in healthy normal controls, cohorts with essential hypertension stratified by renin status, and patients with secondary hypertension.
  • The assay achieved a detection sensitivity of 0.1 pg/tube with 100% recovery, displaying 100% cross-reactivity with angiotensin III and less than 0.1% with angiotensin I.
  • Fasting plasma angiotensin II in normal subjects averaged 12.0 ± 2.1 pg/ml (mean ± SE), with a reference range of 3.0 to 21.3 pg/ml.
  • Patients with secondary hypertension exhibited lower angiotensin II levels in primary aldosteronism and higher levels in renovascular hypertension compared to healthy controls.

Abstract

A very sensitive and simplified direct radioimmunoassay system for plasma angiotensin II was developed using the antiserum against synthetic angiotensin II (final dilution = 1: 1,500,000) in combination with 125I-labeled angiotensin II (specific activity = 1,600 microCi/micrograms). In this assay system, it was possible to carry out a direct assay using 100 microliter of plasma without any extraction procedure. This conclusion was supported by 100% recovery, parallelism of plasma samples against the standard curve, and no difference in hormone levels, there was also a high positive correlation between the plasma angiotensin II levels measured by this direct assay and the dowex column extraction method. The sensitivity of this assay system was 0.1 pg/tube, which is the highest sensitivity in studies reported to date. The cross-reactivities of angiotensin III and I against this antiserum were 100% and less than 0.1%, respectively, suggesting that the antiserum was very specific for the C-terminal of angiotensin II. Plasma angiotensin II levels in normal subjects after overnight fasting ranged from 3.0 to 21.3 pg/ml (12.0 +/- 2.1 pg/ml, mean +/- SE). By comparison, plasma angiotensin II levels of patients with essential hypertension were similar to those in the normal renin group, lower than those in low renin group and higher than those in high renin group. In patients with secondary hypertension, levels were lower in those with primary aldosteronism and higher in those with renovascular hypertension when compared to normal subjects.