The potential clinical usefulness of electron density (ED) imaging, that can be directly estimated using dual-layer spectral computed tomography (CT), has been poorly investigated. We explored whether ED imaging might improve thrombus identification compared to conventional imaging in vitro. We evaluated mechanical thrombectomy material obtained from patients with acute ischemic stroke (AIS) treated in a tertiary level stroke center and immediately fixed in 10% neutral buffered formalin and stored in polystyrene test tubes. The test tubes were immersed in a bucket of water for evaluation by spectral CT, along with scattered control tubes. All images were obtained using a dual-layer detector CT scanner. Each tube was assessed using multiparametric side-by-side view of conventional CT (120 kVp), low monoenergetic imaging (40 keV), and ED images. Fifty-eight polystyrene tubes were analyzed, comprising 52 tubes with thrombectomy material of at least 1 mm