January 6, 2026Magnetic Resonance in MedicineOpen Access

Rapid Free‐Breathing and Automated 2D Shimming of the Lung at 3 T

View Full Paper
Ask AI
Bookmark
Share

Authors

PPPavlos PanosGBGrzegorz BaumanOBOliver Bieri

Discussion

Loading...

Member takes

Overview

Automated shimming improves image quality and reduces frequency offsets in lung imaging, indicating potential clinical benefits.

Key Points

  • To develop and validate a rapid method for shimming the lung using a fully automated approach.
  • Used a custom 2D RF spoiled gradient echo sequence for field mapping of the lung.
  • Acquired four echo‐shifted coronal images at 3T with a minimum TE of 0.57 ms and inter‐echo spacing of 0.1 ms.
  • Performed imaging in free‐breathing mode with three repetitions totaling 4.2 s scan time.
  • Shimming of the lung was successfully achieved in all volunteers in under 10 seconds.
  • Subject-specific shimming reduced mean frequency offset in the lung by up to 45 Hz.
  • Improved bSSFP signal homogeneity led to more uniform functional images, enhancing image quality.