Five instruments have been tested for their linearity, sensitivity, accuracy and specificity to actinium-225 quantification using francium-221 and bismuth-213 equilibria. Francium-221 equilibrium was deemed acceptable for TLC RCY% using HPGe and gas-filled proportional counter. Gamma well counting, LSC, and plastic silicon detector required bismuth-213 equilibrium for accurate RCY%. Low content free actinium-225 impurity was accurately reported for all methods except for the silicon detector. Overall, this investigation sheds light on the appropriate analytical methods for 225Ac-radiopharmaceutical QC considering secular equilibrium.
Toro-González et al. (Sat,) studied this question.