Los puntos clave no están disponibles para este artículo en este momento.
Within the limits of present study, the following conclusions were made: (1) WC/CTa screws provided higher preload than noncoated Ta screws in all three implant connection systems. (2) The initial removal torque for Ta screws required higher force than WC/CTa screws, whereas postload removal torque for Ta screws was lower than WC/CTa screws. Calculated Ta screw preload loss percent was higher than for WC/CTa screws, suggesting that WC/CTa screws were more effective in maintaining the preload than Ta screws. (3) Internal conical connections were more effective in maintaining the screw preload in cyclic loads than external-hex butt joint connections.
Park et al. (Fri,) studied this question.