Aims: Photon-counting detector computed tomography (PCD-CT), which allows the exclusion of electronic noise, shows promise for significant dose reduction in coronary CT angiography (CCTA). This study aimed to assess the radiation dose and image quality of CCTA using PCD-CT, combined with high-pitch helical scanning and an ultra-low tube potential of 70 kVp, and investigate the effect of a sharp kernel on image quality and stenosis assessment in such an ultra-low-dose CCTA setting. Methods and results: < 0.001 for both). Conclusion: PCD-CT technology with high-pitch helical scanning and the tube potential of 70 kVp can provide CCTA with ultra-low radiation exposure (DLP, 29 mGy·cm). The noise reduction capability of PCD-CT allows the use of a sharp kernel even in this low-dose CCTA setting without compromising image quality, potentially improving the evaluation of coronary artery stenosis.