PURPOSE: Artifacts caused by strong lipid signals pose challenges in body chemical exchange saturation transfer (CEST) imaging. This study aimed to develop an accurate water-fat reconstruction method based on the multi-echo Dixon technique to remove lipid artifacts in CEST imaging. THEORY AND METHODS: mapping. The proposed method was compared with the reported 3-point Dixon method and the fixed multi-peak model in a phantom study using a fat-free Z-spectrum obtained from MR spectroscopy acquisition as the ground truth. This method was also validated by in vivo experiments on human breast. RESULTS: In the phantom experiments, the Z-spectrum from the SMPM-based method agreed well with the fat-free Z-spectrum from CEST-PRESS (point-resolved spectroscopy), validating the effective removal of lipid artifacts, while a decrease or a rise that appeared at -3.5 ppm was observed in the Z-spectrum from the 3-point method and the FMPM-based method, respectively. In the in vivo experiments, no lipid artifacts were observed in the Z-spectrum or the amide CEST map from the SMPM-based method in the fibro-glandular region of the breast with high fat fractions. CONCLUSION: The SMPM-based method successfully removes lipid artifacts and significantly improves the accuracy of CEST contrast.
Zhao et al. (Wed,) studied this question.