Cable-driven minimally invasive surgical robots suffer from significant motion inaccuracies due to nonlinear transmission effects such as friction, elasticity, and hysteresis. These factors lead to strong nonlinear and direction-dependent behaviors, making accurate modeling and compensation challenging. To address this issue, this study investigates the error characteristics of a cable-driven surgical robot prototype based on its structural features. A kinematic model is first established, and geometric errors are corrected through Denavit–Hartenberg (DH) parameter identification using a least-squares method. To further characterize nonlinear effects, the LuGre friction model and equivalent stiffness theory are introduced to analyze friction and cable deformation behaviors. Since physics-based models alone cannot accurately capture the coupled nonlinear errors, a radial basis function (RBF) neural network is employed to approximate the residual errors. To enable real-time implementation, the predicted errors are further simplified using equivalent polynomial functions for efficient compensation. Experimental results demonstrate that the proposed method significantly improves the motion accuracy of the cable-driven system, effectively reducing both tracking error and hysteresis effects. By integrating mechanism-based modeling with data-driven compensation, this approach provides a practical and effective solution for precision enhancement in cable-driven surgical robotic systems.
Yu et al. (Mon,) studied this question.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: