July 14, 2022Scandinavian Cardiovascular Journal13 citations

Fast acquisition of left and right ventricular function parameters applying cardiovascular magnetic resonance in clinical routine – validation of a 2-shot compressed sensing cine sequence

View Full Paper
JGJan GröschelCAClemens AmmannLZLeonora Zange

Key Result

The 2-shot compressed sensing cine sequence significantly reduced mean scan time compared to conventional cine (178 ± 36 s vs 313 ± 65 s, p < 0.001) while yielding equivalent cardiac morphology.

Study Design

Type

Cross-Sectional (n=155)

Structured PICO

Does a 2-shot compressed sensing cine sequence yield equivalent cardiac morphology in less time compared to conventional cine in patients undergoing CMR?

P
Population
155 consecutive patients with various cardiac diseases scanned during clinical routine
I
Intervention
Prototype 2-shot compressed sensing (CS) cine sequence on a 1.5 T CMR
C
Comparator
Conventional cine sequence on a 1.5 T CMR
O
Outcome
LV and RV morphology (function parameters), subjective and objective image quality, and scan timessurrogate

A 2-shot compressed sensing cine sequence significantly reduces CMR scan time while maintaining equivalent image quality and cardiac function parameter accuracy compared to conventional sequences.

Main Result

Absolute Event Rate: 178% vs 313%

p-value: p=< .001

Abstract

Objectives. To evaluate if cine sequences accelerated by compressed sensing (CS) are feasible in clinical routine and yield equivalent cardiac morphology in less time. Design. We evaluated 155 consecutive patients with various cardiac diseases scanned during our clinical routine. LV and RV short axis (SAX) cine images were acquired by conventional and prototype 2-shot CS sequences on a 1.5 T CMR. The 2-shot prototype captures the entire heart over a period of 3 beats making the acquisition potentially even faster. Both scans were performed with identical slice parameters and positions. We compared LV and RV morphology with Bland-Altmann plots and weighted the results in relation to pre-defined tolerance intervals. Subjective and objective image quality was evaluated using a 4-point score and adapted standardized criteria. Scan times were evaluated for each sequence. Results. In total, no acquisitions were lost due to non-diagnostic image quality in the subjective image score. Objective image quality analysis showed no statistically significant differences. The scan time of the CS cines was significantly shorter (p < .001) with mean scan times of 178 ± 36 s compared to 313 ± 65 s for the conventional cine. All cardiac function parameters showed excellent correlation (r 0.978–0.996). Both sequences were considered equivalent for the assessment of LV and RV morphology. Conclusions. The 2-shot CS SAX cines can be used in clinical routine to acquire cardiac morphology in less time compared to the conventional method, with no total loss of acquisitions due to nondiagnostic quality.Trial registration ISRCTN12344380. Registered 20 November 2020, retrospectively registered.

Ask AI
Helpful
Bookmark
Share
View Full Paper

Cite This Study

Gröschel et al. (2022) conducted a cross-sectional in various cardiac diseases (n=155). 2-shot compressed sensing (CS) cine sequence vs. Conventional cine sequence was evaluated on Scan time (p=< .001). The 2-shot compressed sensing cine sequence significantly reduced mean scan time compared to conventional cine (178 ± 36 s vs 313 ± 65 s, p < 0.001) while yielding equivalent cardiac morphology.

synapsesocial.com/papers/6a16c6607cba52b0f77b9493https://doi.org/10.1080/14017431.2022.2099010
Ask AI
Helpful
Bookmark
Share
View Full Paper

Also Consider

Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context:

  1. 1Assessment of myocardium at risk with contrast enhanced steady-state free precession cine cardiovascular magnetic resonance compared to single-photon emission computed tomography2010 · 76 citations
  2. 2Cardiac MR imaging: current status and future direction.2015 · 116 citations
  3. 3Standardized cardiovascular magnetic resonance imaging (CMR) protocols: 2020 update2020 · 1,246 citations
  4. 460-S Retrogated Compressed Sensing 2D Cine of the Heart: Sharper Borders and Accurate Quantification2021 · 4 citations
  5. 5Compressed sensing acceleration of cardiac cine imaging allows reliable and reproducible assessment of volumetric and functional parameters of the left and right atrium2021 · 15 citations