PulseExploreJournal ClubResearchersJournals
Instagram
HomeJournal ClubExplore
Synapse
⌘+K
Synapse
February 1, 1996Anesthesia & Analgesia

Oscillometric Blood Pressure Measurements by Different Devices Are Not Interchangeable

View Full Paper
Ask AI
Bookmark
Share

Why the study?

Are blood pressure measurements obtained by different oscillometric devices (SpaceLabs, Marquette, Dinamap) interchangeable?

Population

25 patients with major depression treated with electroconvulsive therapy (ECT)

Comparison

Blood pressure measurement by SpaceLabs or… vs Simultaneous blood pressure measurement on the…

Design

Cross-sectional

Follow-up

up to 5-7 min after ECT

Key result

Systolic blood pressure differences exceeded 10 mm Hg in 31% and 32% of comparisons for SpaceLabs and Marquette devices versus Dinamap, respectively, indicating they are not interchangeable.

Authors

MKMark KaufmannHPHans ParggerLDLambertus J. Drop

Discussion

Loading...

Member takes

Overview

BP readings from different oscillometric devices may vary substantially; leaves open whether device standardization is needed in hypertension research.

Study Design

Type

Observational (n=25)

Structured PICO

Are blood pressure measurements obtained by different oscillometric devices (SpaceLabs, Marquette, Dinamap) interchangeable?

P
Population
25 patients with major depression treated with electroconvulsive therapy undergoing simultaneous blood pressure measurements with different oscillometric devices.
E
Exposure
Blood pressure measurement by SpaceLabs (12 patients, 182 data points) or Marquette (13 patients, 193 data points) oscillometric devices
C
Comparator
Simultaneous blood pressure measurement on the opposite arm by Dinamap 1846SX (DIN)
O
Outcome
Agreement and concordance of systolic and diastolic blood pressure measurements between devicessurrogate

Blood pressure measurements obtained by different brands of oscillometric devices show significant variability and should not be considered interchangeable in clinical practice.

Cite This Study

Kaufmann et al. (1996) conducted an observational in Major depression treated with electroconvulsive therapy (n=25). SpaceLabs and Marquette oscillometric blood pressure devices vs. Dinamap 1846SX device was evaluated on Standard deviation of the difference and mean difference in blood pressure measurements. Systolic blood pressure differences exceeded 10 mm Hg in 31% and 32% of comparisons for SpaceLabs and Marquette devices versus Dinamap, respectively, indicating they are not interchangeable.

synapsesocial.com/papers/6a1fe94275fc4a116b2e4b4bhttps://doi.org/10.1097/00000539-199602000-00029
View Full Paper
Ask AI
Bookmark
Share