Two Anger scintillation cameras have been combined into a positron imaging system. Count rate capability has been extended by using absorbers to reduce scattered radiation from the object, by using 1" thick crystals to improve detection efficiency for annihilation radiation, and by implementing pulse shortening techniques and low dead time D. C. coupled electronics. The main characteristics of the device in its present form are high sensitivity (200 cts/sec/μCi), high resolution (system resolution less than 10 mm FWHM) and useful clinical count rates up to 8000 cts/sec.