August 24, 20051 citations

A Cardiac Monophasic Action Potential Recording Array

View Full Paper
FWFrancis X. WitkowskiCentrum Badań JakościPPPatricia A. PenkoskeWashington University in St. Louis

Key Result

A technique using sintered Ag/AgCl button electrodes with controlled suction provided stable in vivo monophasic action potential recordings for up to 4 hours.

Structured PICO

P
Population
In vivo models for cardiac electrophysiology
I
Intervention
Sintered Ag/AgCl button monophasic action potential (MAP) electrodes held by a nylon mesh sock with controlled suction
O
Outcome
Stability and reproducibility of monophasic action potential (MAP) recordingssurrogate

A novel sintered Ag/AgCl electrode array with suction provides stable, reproducible in vivo monophasic action potential recordings during defibrillatory shocks.

Abstract

In vivo intracellular cardiac recordings using conventional glass microelectrodes to study the mechanisms of action of defibrillatory shocks are limited by electrode fragility. A technique was devised using sintered Ag/AgCl electrodes that allows multiple simultaneous monophasic action potential (MAP) recordings in vivo. Button MAP electrodes were designed that are heltonto the epicardial surface by a nylon mesh sock, with controlled application of suction used to obtain the MAP waveform. External vacuum application as well as epicardial fixation were found useful in studies of defibrillatory shocks to minimize shock hazards as well as optimize MAP reproducibility and stability. Stable recordings were obtained for 30 sec intervals every 3 minutes for periods up to 4 hours from the same epicardial site. This technique may prove useful in the in vivo determination of instantaneous activation and shock induced repolarization alterations.

Ask AI
Helpful
Bookmark
Share
View Full Paper

Cite This Study

Witkowski et al. (2005) studied Cardiac electrophysiology. Sintered Ag/AgCl button MAP electrodes with controlled suction was evaluated on Stable monophasic action potential (MAP) recordings. A technique using sintered Ag/AgCl button electrodes with controlled suction provided stable in vivo monophasic action potential recordings for up to 4 hours.

synapsesocial.com/papers/6a224b4000d082f62f972b23https://doi.org/10.1109/iembs.1990.691232
Ask AI
Helpful
Bookmark
Share
View Full Paper

Also Consider

Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context:

  1. 1A simultaneous multichannel monophasic action potential electrode array for in vivo epicardial repolarization mapping2001 · 18 citations
  2. 2Modifications of electrode design for recording monophasic action potentials in animals and humans1987 · 11 citations
  3. 3Recording monophasic action potentials using a platinum-electrode ablation catheter2000 · 10 citations
  4. 4Transmural recording of monophasic action potentials2002 · 10 citations
  5. 5In vitro validation of a new cardiac catheter technique for recording monophasic action potentials1986 · 177 citations