The YuWell YE680B automated upper-arm BP monitor met ISO accuracy requirements with a mean difference of -0.3 ± 6.77 mmHg for SBP and 0.4 ± 6.08 mmHg for DBP compared to a mercury sphygmomanometer.
Cross-Sectional (n=100)
Does the YuWell YE680B automated upper-arm BP monitor accurately measure blood pressure compared to a mercury sphygmomanometer in adults and children aged ≥3 years?
The YuWell YE680B automated upper-arm BP monitor meets ISO accuracy requirements for clinical and home use in adults and children aged ≥3 years.
Mean Difference: -0.3
Accurate blood pressure (BP) measurement is essential for hypertension management. Automated oscillometric BP monitors must be validated according to international standards before clinical and home use. This study aimed to evaluate the accuracy of the YuWell YE680B automated upper-arm BP monitor in accordance with the ISO 81060-2:2018 Universal Standard and its amendments, Amd.1:2020 and Amd.2:2024. A validation study was conducted using the same-arm sequential measurement method in a mixed population of adults and children aged greater than or equal to 3 years. Reference BP measurements were obtained by two trained observers using a calibrated mercury sphygmomanometer and a dual-head stethoscope. Device accuracy was assessed according to ISO criteria 1 and 2, and agreement was evaluated using Bland-Altman analysis. A total of 100 participants contributed to 300 valid paired measurements. The mean difference (test - reference) was -0.3 ± 6.77 mmHg for SBP and 0.4 ± 6.08 mmHg for DBP, fulfilling the requirements of ISO criterion 1. The SD of the participant-level mean differences was 5.81 mmHg for SBP and 5.32 mmHg for DBP, meeting the thresholds of ISO criterion 2. Bland-Altman analysis demonstrated good agreement across the measured BP range, with differences evenly distributed across the supported arm circumference range (17-52 cm). The YuWell YE680B automated upper-arm BP monitor met the accuracy requirements of ISO 81060-2:2018/Amd.1:2020/Amd.2:2024 in a mixed population of adults and children aged greater than or equal to 3 years. These findings support its use for BP measurement in clinical and home settings.
Liang et al. (Wed,) conducted a cross-sectional in Blood pressure measurement (n=100). YuWell YE680B automated upper-arm BP monitor vs. Calibrated mercury sphygmomanometer was evaluated on Mean difference (test - reference) for SBP and DBP (MD -0.3 ± 6.77 mmHg for SBP and 0.4 ± 6.08 mmHg for DBP). The YuWell YE680B automated upper-arm BP monitor met ISO accuracy requirements with a mean difference of -0.3 ± 6.77 mmHg for SBP and 0.4 ± 6.08 mmHg for DBP compared to a mercury sphygmomanometer.