PURPOSE: To compare and validate the accuracy and ease of use of handheld autorefractors against retinoscopic refraction by an ophthalmologist for assessing the visual acuity of older adults in India. METHODS: 190 patients were enrolled at the Sankara Eye Hospital in Bangalore, India, to undergo refraction using three different handheld devices-Retinomax (Nikon Inc., Japan), Netra (Eyenetra, Inc., USA), and QuickSee (PlenOptika, Inc., USA)-and the results were compared with cycloplegic retinoscopy and refraction done by an ophthalmologist. We analyzed the mean, standard deviation (S.D.), and Bland-Altman comparison of dioptric (D) power accuracy. RESULTS: The difference between the handheld devices and subjective refraction for each device was: Retinomax (N = 186), mean -0.41 D, S.D. 2.14; Netra (N = 179), mean 0.61 D, S.D. 2.20; and QuickSee (N = 182), mean -0.05 D, S.D. 1.04. CONCLUSION: The QuickSee and the Retinomax may be used successfully as refraction screening tools in epidemiologic studies of adults in India and as diagnostic tools in low-resource settings.