August 1, 1997Pacing and Clinical Electrophysiology137 citations

Risk Factors for Venous Obstruction in Children with Transvenous Pacing Leads

View Full Paper
FFFelicia H. FigaBMBrian W. McCrindleJBJean‐Luc Bigras

Key Result

A lead cross-sectional area indexed to body surface area > 6.6 mm2/m2 predicted venous obstruction in children with transvenous pacing leads with 90% sensitivity and 84% specificity (P<0.0002).

Study Design

Type

Observational (n=63)

Structured PICO

What are the risk factors for venous obstruction in children with transvenous pacing leads?

P
Population
63 children with transvenous pacing leads evaluated for incidence and risk factors of venous obstruction.
E
Exposure
Transvenous pacing leads
O
Outcome
Incidence and risk factors for venous obstruction (evaluated clinically, echocardiographically, and by venography)safety

In children with transvenous pacing leads, a lead cross-sectional area indexed to body surface area > 6.6 mm2/m2 is a strong predictor of venous obstruction, highlighting the importance of appropriate lead sizing.

Main Result

p-value: p=<0.0002

Abstract

To determine the incidence and risk factors for venous obstruction in children with transvenous pacing leads, 63 children were evaluated clinically and echocardiographically. Patients with abnormal clinical and/or echocardiographic findings were further investigated by venography. Thirteen patients (21%) had evidence of venous obstruction. Venography in 11 (2 refused) showed that severity of obstruction (as defined by percentage of luminal narrowing) was complete (100%) in 3, severe (> 90%) in 4, and moderate (60%-90%) in 5 (1 patient having 2 sites of obstruction). Risk factors for obstruction in 55 patients with single implantation procedures (10 with obstruction; 18%) were sought. Total cross-sectional area of lead(s) was indexed to body surface area at implantation (INDEX). Patients with obstruction had a higher mean INDEX (7.6 +/- 1.6 mm2/m2) than patients without obstruction (4.9 +/- 2.0 mm2/m2); P 6.6 mm2/m2 to best predict obstruction, with a sensitivity of 90% and specificity of 84%. Since pacing is lifelong, sizing of transvenous leads to the child is important to prevent obstruction and preserve venous access.

Ask AI
Helpful
Bookmark
Share
View Full Paper

Cite This Study

Figa et al. (1997) conducted an observational in Children with transvenous pacing leads (n=63). Total cross-sectional area of lead(s) indexed to body surface area (INDEX) vs. Lower INDEX was evaluated on Venous obstruction (p=<0.0002). A lead cross-sectional area indexed to body surface area > 6.6 mm2/m2 predicted venous obstruction in children with transvenous pacing leads with 90% sensitivity and 84% specificity (P<0.0002).

synapsesocial.com/papers/6a4af206f6034ba310a870c3https://doi.org/10.1111/j.1540-8159.1997.tb03594.x