Long-Term Performance of Right Ventricular Implantable Cardioverter-Defibrillator Leads in Arrhythmogenic Right Ventricular Cardiomyopathy and Hypertrophic Cardiomyopathy

YSYuichiro SagawaYNYasutoshi NagataTYTetsuo Yamaguchi

Key Result

In patients with ICDs, the decrease in R-wave amplitude at 5 years was significantly greater in those with ARVC (-3.3 mV) compared to controls without structural heart disease (1.3 mV).

Study Design

Type

Observational (n=50)

Multicenter

No

Structured PICO

Does right ventricular apex ICD lead implantation maintain stable intracardiac parameters over 5 years in patients with ARVC and HCM compared to controls without structural heart disease?

P
Population
50 patients with ICD leads implanted in the right ventricular apex (7 with ARVC, 14 with HCM, and 29 controls without structural heart disease) followed for 5 years.
E
Exposure
Implantable cardioverter-defibrillator (ICD) lead placement in the right ventricular apex with long-term follow-up (5 years) in patients with ARVC or HCM.
C
Comparator
Control patients without structural heart disease (Brugada syndrome, long QT syndrome, idiopathic ventricular fibrillation/tachycardia) with ICD leads in the right ventricular apex.
O
Outcome
Change in ICD-lead intracardiac parameters (R-wave amplitude, pacing threshold, and pacing impedance) between the time of implantation and year 5.surrogate

In patients with ARVC, right ventricular apex ICD leads show significant deterioration in R-wave amplitude over 5 years compared to controls, suggesting the RV septum may be a preferable implantation site.

Main Result

Absolute Event Rate: -3.3% vs 1.3%

p-value: p=0.012

Limitations

  • Retrospective observational study from a single center
  • Small number of patients with ARVC and HCM
  • 5-year observation period may be inadequate to observe differences in pacing threshold and impedance
  • Diagnosis of ARVC was based on Task Force Criteria and gene examination was rarely performed
  • Observation period of 5 years may have been inadequate for a significant difference in pacing threshold and impedance to occur

Abstract

Patients with arrhythmogenic right ventricular cardiomyopathy (ARVC) and hypertrophic cardiomyopathy (HCM) implanted with implantable cardioverter-defibrillators (ICDs) may show a large decrease in R-wave amplitude during long-term follow-up. However, it is unclear whether this decrease is higher in these patients than in those without structural heart disease. This study investigated ICD-lead intracardiac parameters over a long duration in patients with ARVC and HCM and compared these parameters with those of a control group. We included 50 patients (mean age, 55.2 ± 17.2 years; 26% female) with ICD leads in the right ventricular apex, and compared 7 ARVC and 14 HCM patients with 29 control patients without structural heart disease. ICD-lead parameters, including R-wave amplitude, pacing threshold, and impedance during follow-up, were compared. The difference in these parameters between the time of implantation and year 5 were also compared. There were no significant differences in R-wave amplitude at implantation among the 3 groups. The change in R-wave amplitude between the time of implantation and year 5 was significantly greater in the ARVC group (-3.3 ± 5.4 mV, P = 0.012) in comparison to the control group (1.3 ± 2.8 mV); the HCM group showed no significant difference (-0.4 ± 2.3 mV, P = 0.06). Thus, in the ARVC group, R-wave amplitude at year 5 was significantly lower than that in the control group (5.7 ± 4.8 mV versus 12.5 ± 4.5 mV, P = 0.001). In ARVC patients with ICDs, ventricular sensing is likely to deteriorate during long-term follow-up; however, in HCM patients, sensing may not deteriorate.

Ask AI
Helpful
Bookmark
Share
View Full Paper

Cite This Study

Sagawa et al. (2020) conducted an observational in Arrhythmogenic right ventricular cardiomyopathy and hypertrophic cardiomyopathy (n=50). Arrhythmogenic right ventricular cardiomyopathy (ARVC) vs. No structural heart disease was evaluated on Change in R-wave amplitude between implantation and year 5 (mV) (p=0.012). In patients with ICDs, the decrease in R-wave amplitude at 5 years was significantly greater in those with ARVC (-3.3 mV) compared to controls without structural heart disease (1.3 mV).

synapsesocial.com/papers/6a7b547177336a1bae0326dfhttps://doi.org/10.1536/ihj.19-279
Ask AI
Helpful
Bookmark
Share
View Full Paper

Also Consider

Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context:

  1. 1Trends in implantable cardioverter defibrillator and cardiac resynchronisation therapy lead parameters for patients with arrhythmogenic and dilated cardiomyopathies2018 · 4 citations
  2. 2Early and late changes in right heart echocardiographic parameters in hypertrophic cardiomyopathy patients after transvenous cardiac device implantation.2025
  3. 3Implantable Cardioverter/Defibrillator Therapy in Arrhythmogenic Right Ventricular Cardiomyopathy2004 · 293 citations
  4. 4Decrease in Amplitude of Intracardiac Ventricular Electrogram and Inappropriate Therapy in Patients With an Implantable Cardioverter Defibrillator2006 · 30 citations
  5. 5Relationship between cardiac autonomic function and sustained ventricular tachyarrhythmias in patients with an implantable cardioverter defibrillators2010 · 27 citations