In Curie Point pyrolysis gas chromatography a ferromagnetic conductor in contact with the sample is heated inside a low volume glass capillary which is inserted into an rf coil with the carrier gas flowing through the capillary. The intrinsic properties of the conductor permit a self-control of the temperature to the Curie Point. The optimum wire diameter for a rapid warm-up is a function of the rf frequency. In order to have a fast temperature drop after cutting the rf field, small wire diameters are preferable. For such conductors relatively high frequencies are inherently necessary for a fast stabilization to the Curie Point. So far no differences in the self-controlled end-temperatures have been detected for wires of different origin.
Buhler et al. (1970) studied this question.