Synapse
July 1, 1976Medical Physics221 citations

The line spread function and modulation transfer function of a computed tomographic scanner

View Full Paper
PJPhilip F. Judy

Key Points

Key points are not available for this paper at this time.

Abstract

A simple method to measure the line spread function of computed tomographic scanners has been developed. The line spread function of the EMI‐Scanner was measured and the MTF of the scanner was calculated. The reproducibility of the method indicated this technique could be utilized to compare the spatial resolution of various computed tomographic scanners.

Ask AI
Helpful
Bookmark
Share
View Full Paper

Cite This Study

Philip F. Judy (1976) studied this question.

synapsesocial.com/papers/69d925279a6164e50fa3c254https://doi.org/10.1118/1.594283