PulseExploreJournal ClubDebatesTrendingResearchersJournals
Instagram
HomeExploreJournal ClubTrending
Synapse
⌘+K
Synapse
June 1, 1985Circulation520 citationsOpen Access

Continuous-wave Doppler echocardiographic assessment of severity of calcific aortic stenosis: a simultaneous Doppler-catheter correlative study in 100 adult patients.

View Full Paper
PCPhilip J. CurrieJSJ B SewardGRGuy S. Reeder

Structured PICO

Does continuous-wave Doppler echocardiography accurately predict the transaortic pressure gradient compared to catheterization in adults with calcific aortic stenosis?

P
Population
100 consecutive adults with calcific aortic stenosis (mean age 69, range 50 to 89 years)
I
Intervention
Continuous-wave Doppler echocardiographic assessment of aortic valve pressure gradient
C
Comparator
Simultaneous and nonsimultaneous catheter measurements of pressure gradient
O
Outcome
Correlation between Doppler-determined gradient and catheter-measured gradient (maximum, mean, and peak-to-peak)surrogate

Continuous-wave Doppler echocardiography provides a highly reliable, non-invasive estimation of transaortic pressure gradients in adults with calcific aortic stenosis, correlating strongly with simultaneous catheter measurements.

Abstract

Studies of the correlation of aortic valve gradient determined by continuous-wave Doppler echocardiography and that determined at catheterization have, to date, involved young patients and nonsimultaneous measurements. We therefore obtained simultaneous Doppler echocardiographic and catheter measurements of pressure gradient in 100 consecutive adults (mean age 69, range 50 to 89 years). In 63 patients pressure measurements were obtained with dual-catheter techniques and in 37 they were obtained by withdrawal of the catheter from the left ventricle to the ascending aorta. Forty-six of these patients also underwent an outpatient Doppler study 7 days or less before catheterization. The simultaneous pressure waveforms and Doppler spectral velocity profiles were digitized at 10 msec intervals and maximum, mean, and instantaneous gradients (mm Hg) were derived for each. The correlation between the Doppler-determined gradient and the simultaneously measured maximum catheter gradient was r = .92 (SEE = 15 mm Hg), that between the Doppler-determined and mean catheter gradient was r = .93 (SEE = 10 mm Hg), and that between the Doppler and peak-to-peak catheter gradient was r = .91 (SEE = 14). The correlation between the nonsimultaneously Doppler-determined gradient and the maximum gradient measured by catheter was not as strong (r = .79, SEE = 24). The continuous-wave Doppler echocardiographic velocity profile represents the instantaneous transaortic pressure gradient throughout the cardiac cycle. The best correlation with continuous-wave Doppler-determined gradient was obtained with maximum and mean gradients measured by catheter. Continuous-wave Doppler echocardiography can be used to reliably predict the pressure gradient in adults with calcific aortic stenosis.

Ask AI
Helpful
Bookmark
Share
View Full Paper

Cite This Study

Currie et al. (1985) studied this question.

synapsesocial.com/papers/6a011d9536f190e958b4f023https://doi.org/10.1161/01.cir.71.6.1162
Ask AI
Helpful
Bookmark
Share
View Full Paper