May 27, 2009BMC Cardiovascular Disorders12 citationsOpen Access

Reproducibility of wrist home blood pressure measurement with position sensor and automatic data storage

SUSakir UenRFRolf FimmersMBMiriam Brieger

Key Result

Home blood pressure measurement with a wrist device and position sensor demonstrated significantly higher short-term reproducibility for systolic blood pressure compared to office measurement (SD of mean differences 3.81 vs 8.10 mmHg, p < 0.001).

Study Design

Type

Observational (n=97)

Blinding

Single-blind

Multicenter

No

Structured PICO

Does home blood pressure measurement with a wrist device and position sensor improve reproducibility compared to office BP and 24-h ABPM in hypertensive and normotensive subjects?

P
Population
97 adults (mean age 52, 43% female) with and without hypertension, evaluated over a two-week period to compare the reproducibility of three blood pressure measurement methods.
E
Exposure
Home blood pressure measurement using Omron wrist device HEM-637 IT with position sensor and automatic data storage
C
Comparator
Conventional sphygmomanometric office blood pressure and 24-hour ambulatory blood pressure monitoring (24-h ABPM)
O
Outcome
Reproducibility of blood pressure measurement analyzed by the standard deviation and squared differences of mean individual differencessurrogate

Short-term reproducibility of home blood pressure measurement with a validated wrist device featuring a position sensor is superior to office BP and 24-h ABPM, and correlates well with target organ damage.

Main Result

Absolute Event Rate: 3.81% vs 8.1%

p-value: p=<0.001

Limitations

  • Did not compare wrist device reproducibility directly to an upper arm device.
  • Wrist devices have physiological limits regarding accuracy and require strict adherence to instructions.
  • Small sample size for correlation with left ventricular mass index (n=74).

Abstract

BACKGROUND: Wrist blood pressure (BP) devices have physiological limits with regards to accuracy, therefore they were not preferred for home BP monitoring. However some wrist devices have been successfully validated using established validation protocols. Therefore this study assessed the reproducibility of wrist home BP measurement with position sensor and automatic data storage. METHODS: To compare the reproducibility of three different(BP) measurement methods: 1) office BP, 2) home BP (Omron wrist device HEM- 637 IT with position sensor), 3) 24-hour ambulatory BP(24-h ABPM) (ABPM-04, Meditech, Hun)conventional sphygmomanometric office BP was measured on study days 1 and 7, 24-h ABPM on study days 7 and 14 and home BP between study days 1 and 7 and between study days 8 and 14 in 69 hypertensive and 28 normotensive subjects. The correlation coefficient of each BP measurement method with echocardiographic left ventricular mass index was analyzed. The schedule of home readings was performed according to recently published European Society of Hypertension (ESH)- guidelines. RESULTS: The reproducibility of home BP measurement analyzed by the standard deviation as well as the squared differences of mean individual differences between the respective BP measurements was significantly higher than the reproducibility of office BP (p < 0.001 for systolic and diastolic BP) and the reproducibility of 24-h ABPM (p < 0.001 systolic BP, p = 0.127 diastolic BP). The reproducibility of systolic and diastolic office versus 24-h ABPM was not significantly different (p = 0.80 systolic BP, p = 0.1 diastolic BP). The correlation coefficient of 24-h ABMP (r = 0.52) with left ventricular mass index was significantly higher than with office BP (r = 0.31). The difference between 24-h ABPM and home BP (r = 0.46) was not significant. CONCLUSION: The short-term reproducibility of home BP measurement with the Omron HEM-637 IT wrist device was superior to the reproducibility of office BP and 24- h ABPM measurement. Furthermore, home BP with the wrist device showed similar correlations to targeted organ damage as recently reported for upper arm devices. Although wrist devices have to be used cautious and with defined limitations, the use of validated devices with position sensor according to recently recommended measurement schedules might have the potential to be used for therapy monitoring.

Ask AI
Helpful
Bookmark
Share
View Full Paper

Cite This Study

Uen et al. (2009) conducted an observational in Hypertension (n=97). Home blood pressure measurement with Omron HEM-637 IT wrist device vs. Office blood pressure and 24-hour ambulatory blood pressure monitoring was evaluated on Standard deviation of mean differences between two measurement periods for systolic blood pressure (p=<0.001). Home blood pressure measurement with a wrist device and position sensor demonstrated significantly higher short-term reproducibility for systolic blood pressure compared to office measurement (SD of mean differences 3.81 vs 8.10 mmHg, p < 0.001).