Impact of Doppler guidewire size and flow rates on intravascular velocity profiles

View Full Paper
RJRolf JenniMBMartin BüchiHZHans Jörg Zweifel

Key Result

Velocity profiles within vessels of epicardial coronary artery size are not substantially disturbed by the presence of a 0.014" Doppler guidewire.

Structured PICO

Does a 0.014 inch Doppler guidewire disturb velocity profiles in an in vitro model of epicardial coronary arteries?

P
Population
In vitro model using four serially connected silicone tubes corresponding to epicardial coronary arteries, perfused with discarded human whole blood
I
Intervention
0.014 inch Doppler guidewire insertion with antegrade and retrograde perfusion at flow rates ranging from 1.14-5.88 ml/s
O
Outcome
Velocity profile shape factor (fp) and average peak velocity (APV)surrogate

In an in vitro model, the presence of a 0.014 inch Doppler guidewire does not substantially disturb velocity profiles in vessels sized similarly to epicardial coronary arteries.

Abstract

Coronary blood flow velocity measurements by conventional intravascular catheter-based Doppler devices are well known to be affected by catheter size. Moreover, it is of clinical importance that the assessment of maximum vasodilator capacity, i.e., the coronary reserve, might be considerably affected by a shape change of the velocity profile under hyperemia. Therefore, the present in vitro study aimed to assess the impact of a small-size Doppler guidewire on the velocity profiles interrogated in tubes with diameters corresponding to the epicardial coronary arteries at clinically relevant flow rates. A 0.014" guidewire was inserted into four serially connected silicone tubes of known diameter, which were perfused with discarded human whole blood by means of a roller pump. In order to determine the effect of the Doppler guidewire on the velocity profile antegrade and retrograde perfusion were carried out in each vessel segment. Vm, the true mean velocity, was calculated from the time collected flow divided by the corresponding vessel cross-sectional area. Average peak velocity (APV) measurements were obtained by pulling back the Doppler device across each of the four vessel segments with given flow rates ranging from 1.14-5.88 ml/s in antegrade and retrograde direction. The shape factor of the recorded velocity profile (fp) is defined as vm/APV and is generally assumed to be close to 0.5. Antegrade perfusion: APV = 1.63 vm + 5.35 (R2 = 0.98); fp = 0.001 vm + 0.51. Retrograde perfusion: APV = 1.71 vm + 3.33 (R2 = 0.98); fp = 0.001 vm + 0.52. Due to the constant relationship between APV and vm, velocity profiles within vessels of epicardial coronary artery size are not substantially disturbed by the presence of the Doppler guidewire. The slight, but significant increase of experimental fp with increasing flow is at variance with the theoretically expected fp values.

Ask AI
Helpful
Bookmark
Share
View Full Paper

Cite This Study

Jenni et al. (1998) studied this question. 0.014" Doppler guidewire was evaluated on Impact of Doppler guidewire on velocity profile (shape factor fp). Velocity profiles within vessels of epicardial coronary artery size are not substantially disturbed by the presence of a 0.014" Doppler guidewire.

synapsesocial.com/papers/6a601991c2662a747aa62f6ahttps://doi.org/10.1002/(sici)1097-0304(199809)45:1<96::aid-ccd23>3.0.co;2-i
Ask AI
Helpful
Bookmark
Share
View Full Paper