July 3, 2015Journal of Magnetic Resonance Imaging107 citationsOpen Access

Reproducibility and interobserver variability of systolic blood flow velocity and 3D wall shear stress derived from 4D flow MRI in the healthy aorta

View Full Paper
POPim van OoijAPAlexander L PowellWPWouter V. Potters

Key Result

Systolic velocity and wall shear stress derived from 4D flow MRI were highly reproducible between visits (regional ICC 0.9 for both) and showed low interobserver variability in healthy volunteers.

Study Design

Type

Observational (n=14)

Structured PICO

What is the reproducibility and interobserver variability of systolic blood flow velocity and 3D wall shear stress derived from 4D flow MRI in the healthy aorta?

P
Population
14 healthy controls who underwent test-retest 4D flow MRI examinations separated by 16 ± 3 days to assess reproducibility.
E
Exposure
Noncontrast 4D flow magnetic resonance imaging (MRI) with test-retest separated by 16 ± 3 days
O
Outcome
Reproducibility and interobserver variability of 3D aortic velocity vector fields and wall shear stress (WSS) averaged over five systolic timeframessurrogate

Systolic velocity and wall shear stress derived from 4D flow MRI demonstrate high reproducibility and low interobserver variability in healthy volunteers.

Abstract

PURPOSE: To investigate the reproducibility and interobserver variability of 3D aortic velocity vector fields and wall shear stress (WSS) averaged over five systolic timeframes derived from noncontrast 4D flow magnetic resonance imaging (MRI). MATERIALS AND METHODS: Fourteen controls underwent test-retest 4D flow MRI examinations separated by 16 ± 3 days (resolution = 3.0-3.6 × 2.3-2.6 × 2.5-2.7 mm(3) ; TE/TR/FA = 2.5/4.9 msec/7°; Venc = 150 cm/s). Two observers segmented the aorta, and WSS was calculated for both series of scans and both segmentations. Test-retest and interobserver velocity and WSS vectors were compared on a voxel-by-voxel basis in the aorta and on a regional basis by subdividing the aortas in six segments. RESULTS: Test-retest: voxel-by-voxel Bland-Altman analysis revealed small differences (-0.03/-0.02 m/s/Pa), limits of agreement (LOA) of 0.25 m/s/0.29 Pa, and coefficients of variation (CV) of 20% for velocity/WSS. Voxel-by-voxel orthogonal regression analysis showed moderate agreement (slope: 1.14/1.16, intraclass correlation coefficient ICC: 0.76/0.67 for velocity/WSS). The regional analysis revealed a CV of 9%/8% and ICC of 0.9/0.9 for velocity/WSS. Interobserver: voxel-by-voxel difference for WSS was 0, LOA: 0.17/0.19 Pa, CV: 12/13%, slope: 1.01/1.09, ICC: 0.87/0.85 for test/retest. The CV/ICC for WSS in the regional analysis was 4%/1.0 for test and 3%/1.0 for retest. CONCLUSION: Systolic velocity and WSS derived from 4D flow MRI are reproducible between consecutive visits, with low interobserver variability in healthy volunteers.

Ask AI
Helpful
Bookmark
Share
View Full Paper

Cite This Study

Ooij et al. (2015) conducted an observational in Healthy aorta (n=14). 4D flow MRI vs. Test-retest and interobserver comparison was evaluated on Reproducibility and interobserver variability of 3D aortic velocity vector fields and wall shear stress (WSS). Systolic velocity and wall shear stress derived from 4D flow MRI were highly reproducible between visits (regional ICC 0.9 for both) and showed low interobserver variability in healthy volunteers.

synapsesocial.com/papers/6a6d231778a11c550e07d1f7https://doi.org/10.1002/jmri.24959