PulseExploreJournal ClubDebatesTrendingResearchersJournals
Instagram
HomeExploreJournal ClubTrending
Synapse
⌘+K
Synapse
August 1, 1979Archives of Surgery20 citations

Quantitative Transcutaneous Arterial Velocity Measurements With Doppler Flowmeters

View Full Paper
PSPerry M. Shoor

Key Points

Key points are not available for this paper at this time.

Abstract

A controlled in vitro and in vivo comparative evaluation of four Doppler ultrasonic blood velocimeters was performed to evaluate their ability to measure fluid velocity throughout a wide range of flow rates and velocity profiles, during both steady and pulsatile flow. The instruments tested were the Versatone Bidirectional Doppler M-9, the Sonicaid BV380, the Parks 806 B, and the Delalande DUD 400. All instruments responded linearly to velocities of from 5 to 70 cm/s, and were accurate to within 5% through a wide range of in vitro changes in velocity, pulse frequency, and cross-sectional velocity profile. Quantitative velocity measurements and their derivatives are an inexpensive, simple, and useful tool in the evaluation conditions of patients with peripheral vascular disease, and widespread clinical trials of their value are appropriate at this time.

Ask AI
Helpful
Bookmark
Share
View Full Paper

Cite This Study

Perry M. Shoor (1979) studied this question.

synapsesocial.com/papers/6a90e361647f40dea9c1a4b9https://doi.org/10.1001/archsurg.1979.01370320054009
Ask AI
Helpful
Bookmark
Share
View Full Paper