MR sequence design to account for nonideal gradient performance

View Full Paper
DWDaniel J. WestFGFelix GlangJEJonathan Endres

Key Points

  • The proposed method reduces ghosting artifacts in echo-planar imaging by an average of 77%.
  • Gradient waveforms are optimized by modeling system imperfections, enabling better imaging performance.
  • Utilizing an empirical gradient system transfer function, the method achieves negligible reconstruction errors.
  • This work underscores the potential for enhancing performance even in resource-limited MRI systems.

Abstract

Abstract Purpose MRI systems are traditionally engineered to produce close to idealized performance, enabling a simplified pulse sequence design philosophy. An example of this is control of eddy currents produced by gradient fields; usually these are compensated by pre‐emphasizing demanded waveforms. This process typically happens invisibly to the pulse sequence designer, allowing them to assume achieved gradient waveforms will be as desired. Although convenient, this requires system specifications exposed to the end user to be substantially down‐rated, as pre‐emphasis adds an extra overhead to the waveforms. This strategy is undesirable for lower performance or resource‐limited hardware. Instead, we propose an optimization‐based method to design precompensated gradient waveforms that (i) explicitly respect hardware constraints and (ii) improve imaging performance by correcting k‐space samples directly. Methods Gradient waveforms are numerically optimized by including a model for system imperfections. This is investigated in simulation using an exponential eddy current model, then experimentally using an empirical gradient system transfer function on a 7T MRI system. Results Our proposed method discovers solutions that produce negligible reconstruction errors while satisfying gradient system limits, even when classic pre‐emphasis produces infeasible results. Substantial reduction in ghosting artifacts from echo‐planar imaging was observed, including an average reduction of 77% in ghost amplitude in phantoms. Conclusions This work demonstrates numerical optimization of gradient waveforms, yielding substantially improved image quality when given a model for system imperfections. Although the method as implemented has limited flexibility, it could enable more efficient hardware use and may prove particularly important for maximizing performance of lower‐cost systems.

Ask AI
Helpful
Bookmark
Share
View Full Paper

Cite This Study

West et al. (2025) studied this question.

synapsesocial.com/papers/68d4724731b076d99fa6a8fbhttps://doi.org/10.1002/mrm.70093
Ask AI
Helpful
Bookmark
Share
View Full Paper

Also Consider

Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context:

  1. 1MR sequence design using digital twins of non-idealized hardware2024
  2. 2Overcoming system imperfections using end-to-end MR sequence design.2024
  3. 3Time optimal control‐based RF pulse design under gradient imperfections2019 · 17 citations
  4. 4Bandlimited waveform designs: improving gradient waveform fidelity and controlling acoustic response2025
  5. 5Measuring gradient waveforms effectively with fully compensated variable-prephasing2024