PulseExploreJournal ClubDebatesTrendingResearchersJournals
Instagram
HomeExploreJournal ClubTrending
Synapse
⌘+K
Synapse
August 14, 2024Proceedings on CD-ROM - International Society for Magnetic Resonance in Medicine. Scientific Meeting and Exhibition/Proceedings of the International Society for Magnetic Resonance in Medicine, Scientific Meeting and Exhibition2 citations

MR Signature Matching (Mrsigma) for Volumetric Mri With Less Than 300ms Latency on a 1.5t MR-Linac System

View Full Paper
SSSyed SiddiqVMVíctor MurrayNTNeelam Tyagi

Key Points

Key points are not available for this paper at this time.

Abstract

The MR-Linac system offers a platform to adapt and monitor radiotherapy of tumors affected by continuous motion in real-time. However, real-time MRI technology is restricted to 2D imaging, which limits performance. MR Signature Matching (MRSIGMA) is a recently developed real-time 3D MRI technique with the MR-Linac as a target. However, MRSIGMA was tested only in a non-real-time scenario. This work develops a real-time implementation of MRSIGMA on the Elekta Unity 1.5T MR-Linac, including raw data transmission to an external computer, deep learning 4D reconstruction and correlation-based matching to demonstrate total imaging latency lower than 300ms.

Ask AI
Helpful
Bookmark
Share
View Full Paper

Cite This Study

Siddiq et al. (2024) studied this question.

synapsesocial.com/papers/68e5c453b6db64358755ac62https://doi.org/10.58530/2023/1023
Ask AI
Helpful
Bookmark
Share
View Full Paper