June 1, 1947Journal of the Optical Society of America210 citations

Multiplier Photo-Tube Characteristics: Application to Low Light Levels*

View Full Paper
RER. W. EngstromNorthwestern University

Key Points

Key points are not available for this paper at this time.

Abstract

Measurements in such fields as photometry, spectrometry, and astronomy, dealing with very low light levels, may be assisted by the use of photo-multiplier tubes. Spectral characteristics of visible (S-8), blue (S-4), and ultraviolet (S-5) sensitive multiplier tubes are presented. Data are shown illustrating the wide range of linearity of photo-multiplier types 931-A, 1P21, 1P22, and 1P28. The tubes are linear at currents up to the point where space charge in the output stages reduces the ratio of anode current to cathode light flux. Dark currents and the limitations which they introduce in practice are discussed. Fatigue effects are discussed and data shown which indicate good stability at low current levels. The ultimate sensitivity of these tubes when refrigerated to various degrees has been investigated; with suitable precautions, pulses attributable to individual photoelectrons can be counted.

Ask AI
Helpful
Bookmark
Share
View Full Paper

Cite This Study

R. W. Engstrom (1947) studied this question.

synapsesocial.com/papers/6a177962d990e918e6b41c2bhttps://doi.org/10.1364/josa.37.000420