PulseExploreJournal ClubDebatesTrendingResearchersJournals
Instagram
HomeExploreJournal ClubTrending
Synapse
⌘+K
Synapse
March 1, 1998Journal of Cardiovascular Electrophysiology104 citations

Enhanced Detection Criteria in Implantable Defibrillators

View Full Paper
JBJosép BrugadaLMLluı́s MontMFMARCIO FTGUEIREDO

Structured PICO

Does programming sudden onset and stability detection criteria with a sustained rate duration safety net improve appropriate device management in patients with implantable defibrillators?

P
Population
82 consecutive patients with a Guidant-CPI implantable defibrillator with electrogram storing capability
I
Intervention
Programming sudden onset > 9% and stability < 40 msec in zone 1 of therapy together with a sustained rate duration security mechanism
O
Outcome
Sensitivity and specificity in detecting ventricular tachycardia versus supraventricular arrhythmiassurrogate

Programming sudden onset and stability detection criteria with a sustained rate duration safety net allows for appropriate device management of supraventricular and slow ventricular tachycardias.

Abstract

INTRODUCTION: Enhanced detection criteria in third-generation implantable defibrillators have been implemented to avoid inappropriate therapy of fast supraventricular arrhythmias. We prospectively analyzed the use of these criteria in patients with an implantable defibrillator with electrogram storing capability. METHODS AND RESULTS: In 82 consecutive patients with a Guidant-CPI implantable defibrillator, sudden onset > 9% and stability 9% and stability < 40 msec was the algorithm with the best specificity and sensitivity. CONCLUSION: Programming sudden onset and stability detection criteria with a sustained rate duration safety net for triggering tachycardia therapy results in appropriate device management in most patients with supraventricular and slow (< 210 beats/min) ventricular tachycardias.

Ask AI
Helpful
Bookmark
Share
View Full Paper

Cite This Study

Brugada et al. (1998) studied this question.

synapsesocial.com/papers/6a193e04f2eb401dc788d0d1https://doi.org/10.1111/j.1540-8167.1998.tb00911.x
Ask AI
Helpful
Bookmark
Share
View Full Paper

Also Consider

Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context:

  1. 1Performance of Basic Ventricular Tachycardia Detection Algorithms in Implantable Cardioverter Defibrillators; Implications for Device Programming1997 · 21 citations
  2. 2How to avoid inappropriate therapy2008 · 17 citations
  3. 3Programmable VT Detection Enhancements in Implantable Cardioverter Defibrillator Therapy1995 · 96 citations
  4. 4Safety and Efficacy of a Dual Chamber Implantable Cardioverter Defibrillator Capable of Slow Ventricular Tachycardia Discrimination:2003 · 9 citations
  5. 5Prospective Evaluation of New and Old Criteria to Discriminate Between Supraventricular and Ventricular Tachycardia in Implantable Defibrillators1998 · 45 citations