Safety and efficacy of percutaneous endocardial mapping in sheep heart using a multi-electrode basket-shaped catheter

View Full Paper
EVEnno T. van der VeldeAKA. KarakisPSPaul Steendijk

Key Result

A novel basket-shaped multi-electrode catheter was successfully deployed in 9 sheep within 15 minutes, yielding high-quality electrograms from 29 of 32 electrodes without hemodynamic interference.

PICO

P
Population
9 sheep undergoing left ventricular endocardial mapping using a novel multi-electrode basket catheter.
I
Intervention / Comparator
Basket-shaped multi-electrode catheter
O
Primary Outcome
Safety and efficacy of endocardial mapping (insertion time, hemodynamic interference, electrogram quality)

Abstract

Detailed endocardial catheter mapping to explore cardiac activation patterns is usually carried out with the use of single point catheters. This method suffers from long procedural times and non-optimal resolution due to the limited number of electrodes. In order to localize endocardial electrical activation accurately, multipolar electrode catheter mapping with improved resolution is mandatory. A total of 9 sheep were studied with a newly developed, basket shaped, 8.5 F high-density multipolar catheter consisting of 8 equidistant, collapsible metallic arms, each containing 4 bipolar electrodes. Three of the arms contain radiopaque markers at different levels to determine the spatial orientation of the basket when deployed in the left ventricle. The mapping catheter, integrated with a newly developed computerized mapping system, could be successfully inserted and deployed in all sheep within 15 minutes and without hemodynamic interference. The latter was demonstrated by measuring left ventricular function before and after basket insertion with the use of a conductance catheter. Appropriate electrode-endocardial contact was demonstrated by successful bipolar pacing (2 and 5 mA) in 88% of the electrodes. High quality electrograms were obtained from 29 of the 32 mapping electrodes. Color-coded activation maps were reconstructed during both sinus rhythm and pacing. Postmortem examination of the hearts revealed no catheter-induced trauma. The novel basket shaped multi-electrode catheter facilitates safe, first and accurate endocardial mapping of electrical activity in the left ventricle during sinus rhythm and during pacing in a sheep model.

Ask AI
Helpful
Bookmark
Share
View Full Paper

Cite This Study

Velde et al. (2003) studied Endocardial mapping (n=9). Basket-shaped multi-electrode catheter was evaluated on Safety and efficacy of endocardial mapping (insertion time, hemodynamic interference, electrogram quality). A novel basket-shaped multi-electrode catheter was successfully deployed in 9 sheep within 15 minutes, yielding high-quality electrograms from 29 of 32 electrodes without hemodynamic interference.