Pseudo‐gating: Elimination of periodic motion artifacts in magnetic resonance imaging without gating

View Full Paper
EHE. Mark HaackeGLG. LenzANA.D. Nelson

Key Points

Key points are not available for this paper at this time.

Abstract

This note explains and illustrates a technique of reducing artifacts produced by periodic motion without using physiologic gating. The method is simple and can be applied on any standard MRI unit. For periodic motion, effective gating can be attained by setting the product of the number of acquisitions (at each phase-encoding step), N, times the repeat time, TR, equal to the period of the motion, T. This pseudo-gating can be used with any TR but, if changes in the period occur, is most robust with short TR values. The method is also applicable to multislice and volume imaging. For fast field echo methods, the above rule can be used if the period of the motion is smaller than the total scan time. Otherwise, the scan need only be repeated N times to avoid artifacts and improve signal-to-noise. The method has been implemented to remove both respiratory and/or cardiac motion artifacts.