PulseExploreJournal ClubDebatesTrendingResearchersJournals
Instagram
HomeExploreJournal ClubTrending
Synapse
⌘+K
Synapse
June 1, 2009102 citationsOpen Access

Low-cost differential front-end for Doppler radar vital sign monitoring

View Full Paper
RFRich FletcherJHJing Han

Key Result

A differential front-end design using dual helical antennas reduced background motion noise for heart rate detection at 0.5 meters compared to a traditional single radar unit design.

Structured PICO

I
Intervention
Differential front end design using dual helical antennas and direct-conversion radar circuits operating at 2.460 GHz and 2.510 GHz
C
Comparator
Traditional single radar unit design
O
Outcome
Reduction in background motion noise for heart rate detection at 0.5 meter

A novel differential front-end Doppler radar design reduces background motion noise for non-contact heart rate monitoring.

Abstract

We present a differential front end design for improving the performance of short-range low-cost Doppler radars for vital sign detection with application to automotive driver safety systems, health monitoring, and security screening. By using dual helical antennas each with a 40-degree beamwidth, it is possible to illuminate the body in two adjacent locations to perform a differential measurement. Since only one of the beams illuminates the heart, the baseband signal from the second radar is used for motion cancellation. We demonstrate this approach using a custom-designed dual helical antenna and simple direct-conversion radar circuits implemented in microstrip operating at 2.460 GHz and 2.510 GHz., respectively. Heart rate data is shown for detection distances of 0.5 meter showing reduction in background motion noise compared to data from traditional single radar unit design.

Ask AI
Helpful
Bookmark
Share
View Full Paper

Cite This Study

Fletcher et al. (2009) studied Vital sign detection. Differential front end design using dual helical antennas vs. Traditional single radar unit design was evaluated on Reduction in background motion noise for heart rate detection. A differential front-end design using dual helical antennas reduced background motion noise for heart rate detection at 0.5 meters compared to a traditional single radar unit design.

synapsesocial.com/papers/6a21783f1437d1e11ff1ac9bhttps://doi.org/10.1109/mwsym.2009.5165949
Ask AI
Helpful
Bookmark
Share
View Full Paper