The difference in pulse transit time to the toe and finger measured by photoplethysmography

View Full Paper
MNMeir NitzanBKBoris KhanokhYSYouval Slovik

Key Result

Pulse transit time parameters (E-T PTT and T-F PTTD) decreased as functions of age and systolic blood pressure, but did not significantly depend on diastolic blood pressure.

Study Design

Type

Observational (n=44)

Structured PICO

Does pulse transit time measured by PPG and ECG correlate with age and blood pressure in normotensive males?

P
Population
44 normotensive male subjects evaluated for pulse transit time to the toes and fingers using photoplethysmography and ECG.
E
Exposure
Measurement of pulse transit time (PTT) to toes and fingers using photoplethysmography (PPG) and ECG
O
Outcome
Time delay between ECG R-wave and arrival time of pulses at the toe (E-T PTT), and difference in transit time between toe and finger (T-F PTTD)surrogate

Pulse transit time parameters measured by PPG decrease with age and systolic blood pressure, reflecting structural decreases in arterial compliance.

Abstract

Blood pressure pulse wave velocity (PWV) is a parameter which is related to arterial distensibility. Its direct assessment, by measuring the appearance time of a pressure pulse in two sites along an artery and the distance between the two sites, is complicated and inaccurate. In the current study, pulse transit time (PTT) to the toes and fingers of 44 normotensive male subjects was measured by photoplethysmography (PPG) and ECG. The arrival time of the pulses at the toe and finger was determined from the foot of the systolic rise of the PPG signal, i.e. at end-diastolic time. Two parameters, which are related to PWV, were tested: the time delay between the ECG R-wave and the arrival time of the pulses at the toe (E-T PTT), and the difference in the transit time of the blood pressure pulses between the toe and finger (T-F PTTD). E-T PTT and T-F PTTD decreased as functions of the subject's age and systolic blood pressure (SBP), but their dependence on the diastolic blood pressure (DBP) was not statistically significant. The decrease of the PTT parameters with age is attributed to the direct structural decrease of the arterial compliance with age and not to functional effects associated with the increase of the blood pressure with age, since the PTT parameters did not depend on DBP though the measurements were performed at end-diastole.

Ask AI
Helpful
Bookmark
Share
View Full Paper

Cite This Study

Nitzan et al. (2001) conducted an observational in Normotensive (n=44). Age and systolic blood pressure was evaluated on Time delay between the ECG R-wave and the arrival time of the pulses at the toe (E-T PTT), and the difference in the transit time of the blood pressure pulses between the toe and finger (T-F PTTD). Pulse transit time parameters (E-T PTT and T-F PTTD) decreased as functions of age and systolic blood pressure, but did not significantly depend on diastolic blood pressure.