Detection of Mitral Regurgitation by Doppler Ultrasound

View Full Paper
KKKarina KnutsenEBErling A. BaeESE Sivertssen

Key Result

Pulsed Doppler ultrasound detected mitral regurgitation with 94% sensitivity, 90% specificity, and predictive values of 90% (positive) and 93% (negative).

Study Design

Type

Cross-Sectional (n=64)

Structured PICO

Does pulsed Doppler ultrasound accurately detect mitral regurgitation compared to left ventricular angiography in patients with various heart diseases?

P
Population
64 patients with various heart diseases evaluated for mitral regurgitation using left ventricular angiography and pulsed Doppler ultrasound.
E
Exposure
Pulsed Doppler ultrasound
C
Comparator
Left ventricular angiography (reference standard)
O
Outcome
Detection of mitral regurgitation (sensitivity, specificity, positive and negative predictive values)surrogate

Pulsed Doppler ultrasound demonstrates high sensitivity (94%) and specificity (90%) for the non-invasive detection of mitral regurgitation compared to left ventricular angiography.

Main Result

Effect estimate: Sensitivity 94%, specificity 90%

Limitations

  • Inability to quantitate the regurgitation
  • False positive recordings in patients with coronary heart disease
  • False negative recordings related to problems in placing the ultrasonic sample volume
  • False negatives due to technical difficulty in placing the ultrasonic sample volume
  • False positives possibly due to inconsistent MR from ischemic papillary muscle dysfunction

Abstract

Sixty-four patients with various heart diseases were evaluated for mitral regurgitation (MR) by left ventricular angiography and pulsed Doppler ultrasound within 2-3 days. Adequate Doppler recordings were obtained from 60 patients. The sensitivity of Doppler in detecting MR was 94%, specificity 90% and predictive value for a positive and a negative Doppler examination 90 and 93%, respectively. Two of the 3 false positive Doppler recordings were from patients with coronary heart disease, ischemic papillary muscle dysfunction may have caused MR inconsistently. Two false negative Doppler recordings may be related to problems in placing the ultrasonic sample volume at the systolic retrograde blood stream in the left atrial cavity. It is concluded that pulsed Doppler ultrasound, being a transcutaneous, non-invasive method, is a valuable tool in detecting MR. So far, we have not been able to quantitate the regurgitation.

Ask AI
Helpful
Bookmark
Share
View Full Paper

Cite This Study

Knutsen et al. (1981) conducted a cross-sectional in Mitral regurgitation (n=64). Pulsed Doppler ultrasound vs. Left ventricular angiography was evaluated on Detection of mitral regurgitation (Sensitivity 94%, specificity 90%). Pulsed Doppler ultrasound detected mitral regurgitation with 94% sensitivity, 90% specificity, and predictive values of 90% (positive) and 93% (negative).

synapsesocial.com/papers/6a6cfa8dce524a4339c2a1c3https://doi.org/10.1111/j.0954-6820.1981.tb09829.x
Ask AI
Helpful
Bookmark
Share
View Full Paper