Device Longevity in a Contemporary Cohort of ICD/CRT‐D Patients Undergoing Device Replacement

View Full Paper
FZFrancesco ZanonCMCristian MartignaniEAErnesto Ammendola

Structured PICO

What factors are associated with the service life of ICD and CRT-D devices?

P
Population
953 consecutive patients undergoing ICD or CRT-D replacement across 36 Italian centers between March 2013 and May 2015.
I
Intervention
Implantable cardioverter-defibrillators (ICD) and cardiac resynchronization therapy defibrillators (CRT-D)
C
Comparator
Comparison between device types (single/dual-chamber ICD vs CRT-D) and manufacturers (e.g., Biotronik, Boston Scientific)
O
Outcome
Device longevity (service life) and reasons for replacement

CRT-D devices, higher percentage of ventricular pacing, and specific manufacturers are associated with shorter device longevity, highlighting significant differences in battery life among contemporary devices.

Abstract

INTRODUCTION: The longevity of defibrillators (ICD) is extremely important from both a clinical and economic perspective. We studied the reasons for device replacement, the longevity of removed ICD, and the existence of possible factors associated with shorter service life. METHODS AND RESULTS: Consecutive patients who underwent ICD replacement from March 2013 to May 2015 in 36 Italian centers were included in this analysis. Data on replaced devices were collected. A total of 953 patients were included in this analysis. In 813 (85%) patients the reason for replacement was battery depletion, while 88 (9%) devices were removed for clinical reasons and the remaining 52 because of system failure (i.e., lead or ICD generator failure or a safety advisory indication). The median service life was 5.9 years (25th-75th percentile, 4.9-6.9) for single- and dual-chamber ICD and 4.9 years (25th-75th percentile, 4.0-5.7) for CRT-D. On multivariate analysis, the factors CRT-D device, SC/DC ICD generator from Biotronik, percentage of ventricular pacing, and the occurrence of a system failure were positively associated with a replacement procedure. By contrast, the device from Boston Scientific was an independent protective factor against replacement. Considerable differences were seen in battery duration in both ICD and CRT-D. Specifically, Biotronik devices showed the shortest longevity among ICD and Boston Scientific showed the longest longevity among CRT-D (log-rank test, P < 0.001 for pairwise comparisons). CONCLUSION: Several factors were associated with shorter service life of ICD devices: CRT-D, occurrence of system failure and percentage of ventricular pacing. Our results confirmed significant differences among manufacturers.

Ask AI
Helpful
Bookmark
Share
View Full Paper

Cite This Study

Zanon et al. (2016) studied this question.

synapsesocial.com/papers/6a6f8971e36a167817e13eafhttps://doi.org/10.1111/jce.12990