August 11, 2009Magnetic Resonance in Medicine151 citations

Four‐dimensional (4D) flow of the whole heart and great vessels using real‐time respiratory self‐gating

View Full Paper
SUSergio UribePBPhilipp BeerbaumTSThomas Sangild Sørensen

Structured PICO

Does whole-heart 4D flow imaging using real-time respiratory self-gating provide accurate stroke volume measurements compared to standard 2D phase-contrast scans?

P
Population
15 volunteers and patients with congenital cardiac left-to-right shunting
I
Intervention
Whole-heart 4D flow imaging using real-time respiratory self-gating
C
Comparator
Standard 2D phase-contrast (PC) scans and nongated 4D flow with two averages
O
Outcome
Agreement of stroke volumes (SVs) in the great arteriessurrogate

Real-time respiratory self-gated whole-heart 4D flow imaging provides accurate stroke volume measurements comparable to standard 2D phase-contrast scans, overcoming respiratory motion artifacts.

Abstract

Four-dimensional (4D) flow imaging has been used to study flow patterns and pathophysiology, usually focused on specific thoracic vessels and cardiac chambers. Whole-heart 4D flow at high measurement accuracy covering the entire thoracic cardiovascular system would be desirable to simplify and improve hemodynamic assessment. This has been a challenge because compensation of respiratory motion is difficult to achieve, but it is paramount to limit artifacts and improve accuracy. In this work we propose a self-gating technique for respiratory motion-compensation integrated into a whole-heart 4D flow acquisition that overcomes these challenges. Flow components are measured in all three directions for each pixel over the complete cardiac cycle, and 1D volume projections are obtained at certain time intervals for respiratory gating in real time during the acquisition. The technique was tested in 15 volunteers, in which stroke volumes (SVs) in the great arteries showed excellent agreement with standard 2D phase-contrast (PC) scans. In contrast, nongated 4D flow with two averages had substantial disagreement with 2D flow. Applied to patients with congenital cardiac left-to-right shunting, the precision of flux data was highly beneficial. The methodology presented here has the potential to allow a complete study of flow pathophysiology of the thoracic cardiovascular system from a single free-breathing scan.

Ask AI
Helpful
Bookmark
Share
View Full Paper

Cite This Study

Uribe et al. (2009) studied this question.

synapsesocial.com/papers/6a704aa478a11c550e0a26b4https://doi.org/10.1002/mrm.22090
Ask AI
Helpful
Bookmark
Share
View Full Paper