A digital colorimetric method (ACETimage), which utilizes aldol condensation, crotonization, and resinification, was developed and validated to quantify acetaldehyde in the head fraction of Pisco distillation. The optimal conditions for the reaction were as follows: the head Pisco samples were placed in headspace vials, 20% w/w NaOH was added, and the mixture was boiled in water for 2 min. The Color Grab app was used to capture and analyze images of the reactions, with a screen brightness intensity of 0.5, within a maximum post-reaction time of 10 min. The Euclidean distance (ED-RGB) was the color parameter most sensitive to changes, showing a linear correlation with the square of acetaldehyde concentration, with R2 values ranging from 0.9926 to 0.9976. The limit of detection (LOD) and limit of quantification (LOQ) for the ACETimage method were determined to be 30 and 95.3 mg/L, respectively. A significant correlation was observed between the acetaldehyde content measured using ACETimage and gas chromatography (Spearman’s r = 0.9373). Bland–Altman analysis indicated that the differences between the two methods were within the 95% limits of agreement. ACETimage offers a rapid, cost-effective, and user-friendly solution for monitoring acetaldehyde levels during Pisco distillation, enabling easy implementation in production environments, both artisanal and industrial, with minimal sample preparation and limited personnel training.
Hatta-Sakoda et al. (Thu,) studied this question.