The biological variation and sigma-metrics models were demonstrated to be suitable for setting and evaluating quality targets within and between laboratories. The sigma-metrics model is more flexible, as both the TAE and the risk of failure can be adjusted to the situation-for example, requirements related to diagnosis/monitoring or international authorities. With the aim of reaching (inter)national consensus on advice regarding quality targets for Hb A1c, the Task Force suggests the sigma-metrics model as the model of choice, with default values of 5 mmol/mol (0.46%) for TAE and risk levels of 2σ and 4σ for routine laboratories and laboratories performing clinical trials, respectively. These goals should serve as a starting point for discussion with international stakeholders in the field of diabetes.
Weykamp et al. (Tue,) studied this question.