Do low frequency magnetic fields cause electromagnetic interference and malfunctions in implantable cardioverter-defibrillators?
Low frequency magnetic fields can cause electromagnetic interference and false arrhythmia detections in ICDs at levels below occupational safety limits, highlighting the need for careful risk assessment for employees returning to work.
OBJECTIVES: The aim of this study was to find the electromagnetic interference (EMI) thresholds for several commonly used implantable cardioverter-defibrillators (ICD). DESIGN: Seventeen ICDs were exposed to magnetic fields with different intensities produced by the Helmholtz coil system. Sinusoidal, pulse, ramp, and square-waveforms with a frequency range of 2 Hz to 1 kHz were used. RESULTS: ICD malfunctions occurred in 11 of the 17 ICDs tested. The ICD malfunctions that occurred were false detections of ventricular tachycardia (6/17 ICDs) and ventricular fibrillation (3/17 ICDs), false detection of atrial tachycardia (4/6 dual chamber ICDs) and tachycardia sensing occurring during atrial or ventricular refractory periods (1/17 ICD). In most cases, no interference occurred at magnetic field levels below the occupational safety limits of the International Commission on Non-Ionizing Radiation Protection (ICNIRP). Nevertheless, some frequencies using sine, ramp or square waveforms did interfere with certain ICDs at levels below these limits. No EMI occurred with any of the ICDs below the ICNIRP limits for public exposure. CONCLUSION: Evaluation of EMI should be part of the risk assessment of an employee returning to work after an ICD implantation. The risk assessment should consider magnetic field intensities, frequencies and waveforms.