July 4, 2013Pacing and Clinical Electrophysiology45 citations

Magnetic Resonance Imaging in Patients with Recently Implanted Pacemakers

View Full Paper
HFHannah FriedmanChildren's Hospital ColoradoNANancy AckerMayo ClinicCDConnie DalzellMayo Clinic

Structured PICO

Does MRI scanning cause complications or device malfunction in patients with recently implanted pacemakers compared to those with nonrecently implanted pacemakers?

P
Population
171 patients with implanted pacemakers undergoing 219 MRI scans
I
Intervention
MRI scanning with protocol-driven device reprogramming in patients with recently implanted pacemaker leads (≤42 days)
C
Comparator
MRI scanning with protocol-driven device reprogramming in patients with nonrecently implanted pacemaker leads (>42 days)
O
Outcome
Complications during MRI scanning and clinically significant changes in device function at follow-upsafety

MRI imaging is feasible and safe in patients with recently implanted pacemakers under careful monitoring.

Limitations

  • experience is limited

Abstract

BACKGROUND: Recent studies have shown that magnetic resonance imaging (MRI) of patients with pacemakers can be safely performed under careful monitoring, but they excluded patients with recently implanted devices. Patients with recent implants may be at a greater risk for complications during MRI imaging due to lack of lead and wound maturity. METHODS: We implemented a clinical protocol for MRI imaging of patients with implanted cardiac devices, and prospectively collected data. For this study, we retrospectively analyzed two groups of patients: those with recently implanted (≤42 days) and nonrecently implanted (>42 days) leads at the time of MRI scanning. All devices were interrogated before and after scanning, and were reprogrammed during the scan as per protocol. RESULTS: Of the 219 scans (in 171 patients), eight included patients with recently implanted (range: 7-36 days) and 211 with only nonrecently implanted pacemaker leads. During the scan, there were no complications in the early or late group. In one patient imaged 79 days postimplant, frequent premature ventricular complexes were noted during the scan, requiring no action. No patient reported pain during or immediately after the procedure. No clinically significant changes in function were seen at subsequent follow up (average 104 days post-MRI). Compared to patients with nonrecently implanted leads, there was no difference in any parameter between the two groups. CONCLUSIONS: With a strong clinical indication and with careful monitoring, MRI imaging is feasible in patients with recently implanted pacemakers, although experience is limited.