Systematic Errors in Estimating Mean Blood Pressure from Finger Blood Pressure Measurements

View Full Paper
NONORA E. CHRISTOPHER J. O'CALLAGHAN

Structured PICO

Does calculating mean blood pressure from finger BP measurements accurately reflect measured mean pressure?

P
Population
28 subjects (16 normotensive and 12 hypertensive)
I
Intervention
Calculation of mean blood pressure from finger BP measurements using the formula: diastolic BP + 1/3 pulse pressure
C
Comparator
Directly measured mean blood pressure using the Finapres device
O
Outcome
Difference between calculated and measured mean blood pressure at rest and during vasoactive drug infusions (phenylephrine and isoprenaline)surrogate

Calculating mean blood pressure from finger BP measurements using the standard formula is inaccurate, particularly when pulse wave dimensions are altered by vasoactive drugs.

Abstract

The mean blood pressure (BP) can be accurately estimated from indirect measurements of brachial artery pressure, i.e. mean BP = diastolic BP + 1/3 pulse pressure. Although this equation has been used as a surrogate of mean systemic pressure, it is unknown whether this approximation can be validly applied to distal vascular beds. Therefore we determined the accuracy of this method as an estimate of the mean pressure in distal arteries by measuring finger BP with the Finapres device in 16 normotensive and 12 hypertensive subjects. The "calculated" and measured values of mean BP were compared when subjects were resting and during manoeuvres which aimed to alter the shape of the pulse waveform. Although closely correlated with the measured value, the "calculated" resting mean BP was systematically greater (+2.7+/-0.7 mm Hg, p<0.001). Additionally, the rise in the mean pressure produced by infusion of phenylephrine, an alpha1-adrenoceptor stimulant (16.0+/-1.5 mm Hg) was underestimated by the calculation (13.1+/-1.5 mm Hg, p<0.05). Of even greater concern was that calculating the mean pressure during infusion of isoprenaline (a beta-adrenergic stimulant) suggested the mean pressure had increased by 5.8+/-1.6 mm Hg when it had actually fallen (-2.1+/-2.4 mm Hg, p<0.001 vs. the measured value). Thus, calculating the mean BP from Finapres measurements roughly approximates the measured value when subjects are at rest. However, this estimation becomes inaccurate when pulse wave dimensions are altered, and is probably unsuitable for assessing the acute effects of vasoactive drugs, in particular vasodilators, on BP.